VANE, MRS, CX2305 (USED IN PAIRS)

PART DETAILS

0020-99710
VANE, MRS, CX2305 (USED IN PAIRS)

OEM P/N: 

0020-99710

ESCT P/N: 

ES227-0010

PART TYPE:

2nd

OEM: 

AMAT
MORE INFO

Mass resolving slit vane, CX2305 (OEM P/N 0020-99710), beamline resolving aperture jaw for AMAT ion implanter systems

ESCT P/N ES227-0010, supplied as a matched-pair beamline component for Applied Materials (AMAT) implanter mass-analysis assemblies

The Mass Resolving Slit (MRS) vane, OEM part number 0020-99710 and designated CX2305, is a precision beamline aperture jaw used in the mass-resolution stage of Applied Materials ion implanter platforms. Each vane forms one of the two opposing blades that together define the resolving slit, the narrow gap positioned at the dispersive focal plane of the analysing magnet, where the spatially separated ion species converge after the beam has been bent through the mass-analysis sector. Because the slit is bounded by two facing jaws, the vanes are always supplied and fitted in pairs, hence the “used in pairs” designation in the part name.

In operation, the analysing magnet disperses the extracted ion beam so that ions of differing mass-to-charge ratio (m/z) arrive at slightly different lateral positions in the focal plane. The MRS vanes intercept everything that falls outside the intended aperture width, allowing only the target dopant species (for example a specific boron, phosphorus, arsenic or BF2 charge state) to pass downstream to the acceleration column and wafer. The vanes therefore act simultaneously as a spatial filter and as a beam-defining aperture, and their edge condition, alignment and gap symmetry directly govern the mass resolution the tool can achieve.

As an intercepting component, each vane is continuously bombarded by off-axis ions, neutrals and scattered particles. This makes the MRS vane both a critical optical element and a wear part: its working edges erode and contaminate over time, and worn or mismatched vanes are a common, often under-diagnosed cause of resolution drift and cross-contamination in production implanters.

Functional role of the MRS vane in AMAT ion implanter platforms

The pair of vanes establishes the physical boundary of the resolving aperture and defines the mass window that reaches the wafer. Their combined functions within the beamline include:

  • Defining the resolving slit width at the dispersive focal plane, setting the achievable mass resolution (m/Δm) of the analysing system.
  • Rejecting adjacent and contaminant ion species, including energetic neutrals and unwanted charge states, that fall outside the selected mass window.
  • Intercepting and absorbing the off-axis beam current, dissipating the resulting thermal load away from sensitive downstream optics.
  • Working as a symmetric, opposed pair so the aperture remains centred on the beam axis and the transmitted beam profile stays balanced.
  • Providing a clean, well-defined leading edge so the slit boundary is sharp and the resolution edge is not blurred by erosion or build-up.
  • Preserving beam-current transmission of the wanted species, supporting stable dose rate and predictable wafer throughput.

Material and process performance

MRS vanes sit in a high-vacuum, high-energy environment and must combine dimensional stability with resistance to sputter erosion and thermal stress. The two performance dimensions that matter most in service are described below.

Edge integrity and sputter resistance

The vane edges take the direct impact of intercepted ions and gradually erode through sputtering. Edge geometry and surface condition must be maintained so that the slit boundary stays sharp; a rounded or pitted edge widens the effective transition region of the aperture and degrades mass resolution. Materials and coatings are selected to minimise sputter yield and to limit the generation of secondary particles that could redeposit elsewhere in the beamline.

  • Low sputter-yield working faces to slow edge recession under continuous bombardment.
  • Stable, low-outgassing surfaces that do not compromise beamline vacuum quality.
  • Dimensional tolerance held tightly enough that both jaws of the pair erode and seat consistently.

Thermal load management

Because the vanes absorb the rejected beam power, they experience substantial localised heating, particularly under high-current implant recipes. The component must conduct that heat into its mounting and tolerate repeated thermal cycling without warping, cracking or shifting alignment. Thermal distortion of even one vane changes the slit width and shifts the mass window, so thermal behaviour is integral to resolution stability.

  • Good thermal conduction from the working edge into the mount to limit peak edge temperature.
  • Resistance to thermal fatigue and warping across many implant cycles.
  • Stable mechanical seating so heating does not translate into aperture drift.

Impact on tool performance and yield

Because the MRS vanes define the mass window that reaches the wafer, their condition has a direct line to process results and device yield. Consequences of degraded or mismatched vanes include:

  • Loss of mass resolution, allowing adjacent species or contaminant charge states to leak through the aperture into the wafer.
  • Cross-contamination between dopant recipes when previous-species build-up on the vanes is re-sputtered into the beam.
  • Dose and uniformity drift as the effective slit width changes with edge erosion or thermal distortion.
  • Reduced beam transmission of the target species, lowering achievable dose rate and throughput.
  • Particle excursions and beamline contamination from flaking deposits on worn vane faces.
  • Increased tuning time and beam instability, eroding tool availability and productivity.

Common wear and failure modes

Edge erosion and recession

Continuous ion impact sputters material from the working edges, gradually rounding and receding them. This widens the effective slit transition, softens the resolution edge and shifts the centre of the mass window, the most fundamental wear mode of any resolving aperture jaw and the primary reason for scheduled replacement.

Deposition build-up and re-sputtering

Sputtered and condensed material from earlier implants accumulates on the vane faces. These deposits can flake to generate particles, or be re-sputtered back into the beam, causing memory-effect cross-contamination between species. Build-up also distorts the local geometry of the aperture edge.

Thermal distortion and pair mismatch

High intercepted power can warp a vane or loosen its seating, while unequal wear between the two jaws breaks the symmetry of the slit. A mismatched pair produces an off-centre, asymmetric aperture, which is why the vanes are replaced together rather than singly.

Installation and replacement guidance

Replacement should be carried out by trained implant maintenance personnel during a scheduled preventive-maintenance window, following the OEM service procedure for the specific platform. A typical sequence is:

  • Place the tool in a safe maintenance state, confirm beam off and high voltage disabled, and follow site lockout/tagout before any access.
  • Vent the affected beamline section to atmosphere per the OEM procedure and allow heated components to cool to a safe handling temperature.
  • Open the resolving-slit housing and document the existing vane positions, orientation and shim/alignment settings before removal.
  • Remove both worn vanes as a pair, handling only with clean, powder-free gloves to avoid contaminating the high-vacuum surfaces.
  • Fit the new matched pair, setting slit width and centring per specification and verifying both jaws are symmetric about the beam axis.
  • Reseal the housing, evacuate the beamline and confirm the required base vacuum is re-established before energising.
  • Re-tune the beamline, verify mass resolution and contamination performance, and confirm dose and uniformity against the recipe baseline.

Spare parts strategy for AMAT ion implanter beamlines

Mass resolving slit vanes are recurring consumables tied directly to resolution and contamination performance, so keeping a qualified matched pair on the shelf protects both yield and uptime. ESCT Fab Support helps fabs manage that risk:

  • Components verified to the form, fit and function of OEM P/N 0020-99710, referenced under ESCT P/N ES227-0010.
  • Supplied as matched pairs so both jaws of the resolving slit are replaced together for symmetric wear.
  • A nominal lead time of 2.5 weeks to support planned preventive-maintenance scheduling.
  • A cost-effective alternative to sole-source OEM procurement, reducing spares spend without compromising specification.
  • Knowledgeable support for source and beamline hardware across AMAT implanter platforms.

Frequently asked questions

What does the MRS vane actually do in the implanter?

It forms one jaw of the mass resolving slit at the focal plane of the analysing magnet. Together the pair defines the aperture that lets only the selected dopant species through to the wafer while rejecting adjacent masses, contaminant charge states and neutrals.

Is this an original AMAT part?

No. This is a new, OEM-compatible component supplied by ESCT Fab Support and verified to the form, fit and function of AMAT OEM P/N 0020-99710. It is not an original OEM-branded part and is not sold as such.

Why are the vanes supplied and replaced in pairs?

The resolving slit is bounded by two opposing jaws. If only one is replaced, unequal wear between old and new vanes makes the aperture asymmetric and off-centre, degrading resolution. Fitting a matched pair keeps the slit symmetric about the beam axis.

How do I know when the vanes need replacing?

Typical indicators are a gradual loss of mass resolution, rising cross-contamination or memory effects between recipes, dose and uniformity drift, increased tuning effort, or particle excursions. Visible edge erosion or heavy deposition at a PM inspection also signals end of life.

What is the lead time?

The nominal lead time is approximately 2.5 weeks. Because the vanes are a planned-replacement consumable, holding a matched pair in your spares inventory is the most reliable way to avoid unplanned downtime.

Will fitting these vanes affect my established process?

Components are built to match the dimensional and material characteristics of the OEM part so that, once correctly set and centred, slit width and resolution match your baseline. As with any beamline change, re-tune and verify resolution, contamination, dose and uniformity before returning to production.

Summary

The Mass Resolving Slit vane (CX2305, OEM P/N 0020-99710, ESCT P/N ES227-0010) is a precision beamline aperture jaw that, as one of an opposed pair, defines the resolving slit at the focal plane of an AMAT ion implanter analysing magnet. By setting the mass window and intercepting rejected species, the vanes govern mass resolution, contamination control and beam transmission. As an intercepting consumable they erode, accumulate deposits and can distort thermally, so they are replaced as a matched pair on a planned PM cycle. ESCT Fab Support supplies these vanes verified to OEM form, fit and function, as matched pairs, with a nominal 2.5-week lead time to keep your implanter resolving cleanly and running productively.

Disclaimer: This part is supplied by ESCT Fab Support as a new, OEM-compatible second-source component, verified to the form, fit and function of Applied Materials part number 0020-99710. It is not an original OEM part and is not sold as one. It is not manufactured by, endorsed by, or affiliated with Applied Materials, and all manufacturer names and part numbers are quoted for identification and compatibility purposes only.