Source Extraction Assy Repair

PART DETAILS

0090-91192
Source Extraction Assy Repair

OEM P/N: 

0090-91192

ESCT P/N: 

ES3033-213

PART TYPE:

N/A

OEM: 

AMAT
Repair Offered
MORE INFO

Source extraction assembly repair (OEM P/N 0090-91192), ion source extraction electrode assembly for AMAT Quantum / xR-class ion implantation systems

ESCT P/N ES3033-213, rebuilt and verified source extraction assembly for Applied Materials implanter source modules

The Source Extraction Assembly is the electromechanical structure that sits immediately downstream of the ion source arc chamber on Applied Materials (AMAT) Quantum and xR-class ion implanters. Its job is to convert the dense, near-stationary plasma generated inside the arc chamber into a collimated, high-current ion beam by applying a strong electrostatic field across a set of carefully aligned extraction electrodes. Without a properly functioning extraction assembly, the implanter cannot form a usable beam at all, which makes this component one of the most consequential items in the entire source subsystem.

Mechanically, the assembly comprises the extraction (or “puller”) electrode and suppression electrode, their mounting frame and insulators, the high-voltage feedthroughs, and the precision adjustment hardware that sets the electrode-to-source-aperture gap and beam-line alignment. Because the assembly operates inside a high-vacuum, high-voltage and high-thermal-load environment, directly adjacent to a plasma source running at tens of kilovolts, its electrodes, insulators and alignment surfaces are continuously stressed and eventually require rebuild. This part, P/N 0090-91192, is the repaired and reconditioned extraction assembly supplied to restore that beam-forming function.

ESCT supplies this assembly (ES3033-213) as a fully reconditioned unit so that fab and field-service teams can return an implanter source to full beam current and beam quality without waiting on long OEM lead times or carrying the cost of a brand-new assembly.

Functional role of the source extraction assembly in AMAT ion implanter platforms

In an AMAT implanter source module, the extraction assembly performs the critical first stage of beam transport. It establishes the electric field geometry that draws ions from the plasma meniscus at the source aperture and shapes them into a defined beam before they enter the mass-analysis magnet and the rest of the beamline. The geometry, cleanliness and alignment of this assembly directly determine extracted current, beam emittance and source stability.

  • Applies the high-voltage extraction field that pulls ions from the arc-chamber plasma through the source aperture.
  • Houses the suppression electrode that prevents secondary and back-streaming electrons from flowing into the source and degrading the field.
  • Shapes the emerging beam into a defined cross-section so that downstream optics can transport it efficiently.
  • Maintains a precise, repeatable electrode gap relative to the source aperture, governing perveance-matched current draw.
  • Provides electrical isolation between the source potential and the grounded structure via high-voltage insulators and feedthroughs.
  • Offers the mechanical adjustment datum used to align the extracted beam to the entrance of the analyser magnet.

Electrical and signal performance

High-voltage field integrity

The extraction assembly must sustain extraction potentials of the order of tens of kilovolts across a small vacuum gap without arcing or excessive leakage. Reconditioning restores the dielectric integrity of the insulators, the surface finish of the electrodes and the cleanliness of the high-voltage feedthroughs so that the assembly can hold standoff voltage reliably during conditioning and production.

  • Insulators inspected and cleaned to remove conductive metallisation and tracking paths.
  • Electrode surfaces refinished to suppress field emission and micro-arcing sites.
  • Feedthrough contacts checked for low-resistance, stable high-voltage connection.

Beam current and stability

Extraction efficiency and beam stability depend on the electrode geometry being matched to the source operating point. A correctly rebuilt assembly restores the design gap and alignment so that extracted current responds predictably to source and extraction voltage, keeping dose rate and beam stability within process tolerances.

  • Restored electrode gap returns extracted current to the expected perveance curve.
  • Reduced glitching and arc-down events during high-current and high-energy recipes.
  • Stable suppression voltage that holds back electron back-streaming under load.

Impact on tool performance and yield

Because the extraction assembly defines the very first stage of beam formation, its condition propagates through every downstream metric on the implanter. A degraded assembly costs current, stability and uptime; a sound one underpins consistent dose and angle control.

  • Direct effect on extracted beam current and therefore on achievable throughput.
  • Influence on beam emittance and angular spread, which affects implant uniformity and tilt accuracy.
  • Reduced arc-down and glitch frequency, lowering wafer-level dose errors and rework.
  • Shorter, more predictable source conditioning times after preventive maintenance.
  • Improved source-to-source repeatability across a fleet, supporting tighter process control.

Common failure modes

Electrode erosion and deposition

The extraction and suppression electrodes are continuously bombarded by ions and sputtered source material. Over time the apertures erode and widen while conductive deposits build up on the surrounding surfaces, distorting the field geometry, shifting the perveance match and reducing usable beam current.

Insulator tracking and high-voltage breakdown

Sputtered metal films gradually coat the high-voltage insulators, creating conductive tracking paths that lower standoff voltage. The result is increasing glitch rates, leakage current and eventual flashover, which forces the extraction supply into fault and halts beam delivery until the assembly is cleaned or rebuilt.

Mechanical misalignment and gap drift

Thermal cycling and repeated source changes can loosen mounting hardware and shift electrode alignment, allowing the extraction gap to drift from its set value. This misaligns the beam to the analyser entrance, degrades transmission and produces unstable, hard-to-tune source behaviour.

Installation and replacement guidance

Replacing the source extraction assembly is a high-voltage, high-vacuum maintenance task that should be carried out by trained implant technicians following AMAT service procedures and the site’s lock-out/tag-out policy. The sequence below is a general guide and does not replace the OEM manual.

  • Place the implanter in maintenance mode, disable and discharge the high-voltage extraction supply, and confirm zero energy before access.
  • Vent the source chamber to the appropriate state, then open the source enclosure to expose the extraction assembly.
  • Disconnect the high-voltage feedthroughs and any suppression connections, labelling them for correct reassembly.
  • Release the mounting hardware and withdraw the existing assembly, noting the electrode-gap setting and alignment shims.
  • Fit the reconditioned ES3033-213 assembly, set the extraction gap to the specified value and verify electrode-to-aperture alignment.
  • Reconnect feedthroughs, close the enclosure, pump down to base pressure and confirm the absence of vacuum leaks.
  • High-voltage condition the assembly gradually, then verify extracted current, suppression and beam stability against the source baseline before returning to production.

Spare parts strategy for AMAT ion implanter source modules

The extraction assembly is a high-wear, mission-critical item, so most fabs keep a serviceable spare on the shelf to swap during scheduled source preventive maintenance rather than risk extended downtime. ESCT supports that strategy with reconditioned assemblies that restore form, fit and function at a lower cost and shorter lead time than new OEM supply.

  • Reconditioned to verified form, fit and function for direct drop-in replacement.
  • Typically faster availability than long OEM procurement cycles, reducing tool-down exposure.
  • Lower acquisition cost, freeing budget for additional rotable spares.
  • Supports a swap-and-rebuild rotation that keeps a conditioned assembly always ready.
  • Backed by ESCT inspection and test of high-voltage and mechanical integrity before dispatch.

Frequently asked questions

What does the source extraction assembly actually do?

It applies the high-voltage field that pulls ions out of the arc-chamber plasma and shapes them into a beam, forming the very first stage of beam transport on the implanter. Its geometry and alignment set the extracted current, beam quality and source stability.

Is this an original OEM part?

No. This is not an original Applied Materials part. It is supplied by ESCT as a refurbished and reconditioned assembly that has been verified to the form, fit and function of OEM P/N 0090-91192, intended as a compatible replacement for the corresponding AMAT source extraction assembly.

How do I know when the extraction assembly needs replacing?

Typical indicators include a steady loss of extractable beam current, rising glitch or arc-down rates, increased extraction leakage current, longer conditioning times and a source that becomes difficult to tune. These usually point to electrode erosion or insulator tracking that a rebuild addresses.

Why choose a reconditioned assembly instead of cleaning the existing one?

In-situ cleaning can recover some performance, but eroded electrode apertures and aged insulators cannot be restored by cleaning alone. A reconditioned assembly returns the unit to specified geometry and dielectric integrity, and swapping a ready spare minimises tool-down time compared with bench-rebuilding the removed unit.

Does fitting this part require recalibration?

Yes. After installation the assembly should be high-voltage conditioned and the source re-baselined, with extracted current, suppression voltage and beam alignment verified against the tool’s reference values before returning to production.

Is the assembly compatible with my specific source configuration?

It is built to match the form, fit and function of AMAT P/N 0090-91192. As source modules can vary by tool revision, we recommend confirming your existing assembly part number and source configuration with ESCT before ordering so the correct variant is supplied.

Summary

The Source Extraction Assembly (AMAT P/N 0090-91192, ESCT P/N ES3033-213) is the high-voltage beam-forming structure that draws ions from the arc-chamber plasma and shapes them into the implanter’s primary beam. Its electrode geometry, insulator integrity and alignment govern extracted current, beam quality and source stability, and it is subject to predictable erosion, deposition and high-voltage degradation over service life. Supplied by ESCT as a reconditioned, function-verified unit, it offers fabs a cost-effective, readily available route to restoring full source performance while supporting a solid rotable-spare strategy.

Disclaimer: This part is supplied by ESCT Fab Support as a reconditioned, OEM-compatible component, verified to the form, fit and function of Applied Materials part number 0090-91192. It is not an original OEM part and is not sold as one. It is not manufactured by, endorsed by, or affiliated with Applied Materials, and all manufacturer names and part numbers are quoted for identification and compatibility purposes only.